Instructions for use Fabius Tiro SW 3.n 151
Troubleshooting
Failure of the O2 sensor
Causes for faulty calibration
The calibration was not successful if, after the
calibration of the O
2 sensor, the alarm message
O2 SENSOR FAIL ! is displayed.
Possible causes and remedial measures are
described in the following table.
Cause Remedy
During the calibration, the O
2 sensor was exposed
to an air mixture with extremely high or low oxygen
concentration.
Make sure that the O2 sensor is exposed to ambi-
ent air during the entire calibration.
During the calibration, the O
2 sensor was exposed
to an air mixture with fluctuating oxygen concentra-
tion.
Make sure that the O2 sensor is exposed to ambi-
ent air during the entire calibration.
The O
2 sensor was not exposed to ambient air long
enough before the calibration.
Expose the O2 sensor to ambient air for 2 minutes.
When a new O
2 sensor is connected, expose the
new sensor to ambient air for 15 minutes.
The maximum period of use of the O
2 sensor has
elapsed.
Replace the O2 sensor. Expose the new O2 sensor
to ambient air 15 minutes before calibration.
The O
2 sensor is not connected. Check the O2 sensor. Connect the O2 sensor cor-
rectly and recalibrate.