Instructions for use Fabius Tiro SW 3.n 89
Operation
If the control dial is in position T:
1 Press the 0 key (A) and engage the control dial
(B) to position 0. To ensure pressure
equalization, wait at least 15 seconds.
2 Press the 0 key (A) and set the control dial (B)
counterclockwise to the desired anesthetic gas
concentration.
3 Check the filling level on the sight glass (C)
regularly. The filling level must be between
minimum and maximum.
4 When the refill mark (D) is reached, 250 mL
(normal anesthetic agent bottle) can be refilled.
5 Fill the vaporizer at the latest when the
minimum mark (E) is reached, refer to
instructions for use of the respective vaporizer.
O2 flush
The O2 flush is used for flushing and quickly filling
the breathing system and breathing bag with O2
while bypassing the vaporizer.
During this an unmetered flow of at least 35 L/min
is given to the breathing system and the breathing
bag.
Press the (A) key. O
2 flows for as long as
the key is held down.
Use of the O
2 flush can increase airway pressure
very quickly and abruptly change the gas
concentration.
20960
NOTE
In ManSpont mode, the pressure can increase
quickly and thus trigger the APL valve.