EasyManua.ls Logo

Dräger Fabius Tiro - Connecting the Endotracheal Suction System (Optional)

Dräger Fabius Tiro
224 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Assembly and preparation
70 Instructions for use Fabius Tiro SW 3.n
Mounting the bag elbow
1 Attach the bag elbow (A) on the connection of
the compact breathing system.
2 Attach the breathing bag (B) on the connector
of the bag elbow.
Connecting the endotracheal suction
system (optional)
The endotracheal suction system for Fabius
consists of:
Suction regulator (A)
Suction bottles (D)
1 The suction regulator (A) is attached to a
holder (B). The holder is fastened on the side
GCX rail (C) on Fabius.
2 The suction bottle (D) is attached to a separate
holder.
Suction systems used
When using an ejector suction with drive gas:
Connect the Air connection hose of the suction
system via a 3-way adapter to the gas supply
block or the central gas supply.
For vacuum suction:
Connect the vacuum hose of the suction
system directly to the central gas supply.
Prepare the suction system in accordance with the
associated instructions for use.
21013
A
B
20949
WARNING
Risk of patient injury
Only use the suction system if the ventilation
mode ManSpont is active or the patient is
disconnected from the Y-piece.
A
B
C
D