Instructions for use Fabius Tiro SW 3.n 67
Assembly and preparation
4 Connect the fresh-gas hose (C) to the
corresponding connection of the compact
breathing system.
5 Connect the ventilator hose to the
corresponding connector (D) on the anesthesia
workstation.
6 Connect the other end of the ventilator hose to
the ventilator connector (E) of the compact
breathing system.
If Fabius has a screw connection, the sealing rings
of this screw connection must be undamaged and
clean.
Only hand-tighten the screw connection. Do not
use tools.
Inserting the flow sensor
1 Screw out the expiratory port (A).
2 Remove the flow-sensor guard
*
(B).
3 Insert the flow sensor (C).
4 Replace the flow-sensor guard* (B).
5 Screw the expiratory port (C) back on.
20943
D
E
WARNING
Risk of fire
Residual vapors of easily flammable
disinfectants (e.g., alcohols) and deposits
that were not removed during reprocessing
can ignite when the flow sensor is in use.
Ensure particle-free cleaning and
disinfection.
After disinfection, allow the flow sensor to
air for at least 30 minutes.
Before inserting the flow sensor, check for
visible damage and soiling such as
residual mucus, medication aerosols, and
particles.
Replace flow sensors when damaged,
soiled, or not particle-free.