Technical data
194 Instructions for use Fabius Tiro SW 3.n
Breathing system
Volume with reusable CO2 absorber
(including absorber volume, measured in ManSpont)
Filled, without hoses typically 4000 mL + volume of the breathing bag
Volume with Drägersorb CLIC adapter
(including absorber volume, measured in ManSpont)
Filled, without hoses typically 3700 mL + volume of the breathing bag
Absorber volume
Reusable CO
2 absorber, filled 1500 mL
Disposable CO
2 absorber CLIC absorber Free 1200 mL
Disposable CO
2 absorber CLIC Absorber 800 Free 1200 mL
Compliance
including ventilator hose (without breathing hoses) 0.8 mL/cmH
2O (0.8 mL/hPa)
Flexible arm for breathing bag (optional)
Volume 0.13 L
Compliance 0.13 mL/cmH
2O (0.13 mL/hPa)
Rigid arm for breathing bag (optional)
Volume 0.11 L
Compliance 0.11 mL/cmH
2O (0.11 mL/hPa)
Resistance
In accordance with ISO 80601-2-13, dry, with adult
breathing hose set M30146
1)
Inspiratory: Expira-
tory:
-4.7 cmH2O (-4.7 hPa)
4.4 cmH
2O (4.4 hPa)
In accordance with ISO 80601-2-13, dry, without hos-
es
1)
Inspiratory: Expira-
tory:
-3.7 cmH2O (-3.7 hPa)
3.7 cmH2O (3.7 hPa)
Typical leakage <50 mL/min