Gas still flowing - shut off all gas flow control
valves to activate Sleep Mode.
As soon as the flow control valves are closed, the
message disappears.
3 Close the flow control valves.
Power-saving mode
When Fabius is in Standby mode and there is no
user input for 2.5 minutes, the power-saving mode
is activated. The screen is then replaced by the
screensaver. Press any key to end the
screensaver.
Performing the system test
In Standby mode, a system test can be started.
This test corresponds to the test that is performed
automatically after switching on the anesthesia
workstation. More information can be found in
chapter ''Switching on'' on page 83.
1 Press the Run System Test softkey (A).
– Electrical system components are tested.
– Default settings are restored.
The test results (B) are displayed on the screen.
After completion of the system test, the total test
results (C) are displayed, see chapter ''Checking
the readiness for operation'' on page 85. If the
system test was successful, the Standby mode is
activated.
21975
WARNING
Risk of patient injury
During the system test, the system is
pressurized.
To prevent patient injury, do not perform the
system test on the medical device if a patient
is connected.