Instructions for use Fabius Tiro SW 3.n 153
Troubleshooting
(23/31) APNOEA FLOW !! The apnea flow alarm based
on a time staggering.
In the Volume Control,
Pressure Control, SIMV/PS
modes with Freq ≥ 6 or in
Pressure Support mode
with apnea ventilation deacti-
vated:
Caution = VT <20 mL for >15
seconds
In the ManSpont, SIMV/PS
modes with Freq < 6 or in
Pressure Support mode
with apnea ventilation acti-
vated:
Caution = VT <20 mL for >30
seconds
Breathing/ventilation
stopped.
Check ventilator.
Leakage or disconnection in
the breathing system.
Check breathing system.
Alarm
priority
Alarm Cause Remedy