Instructions for use Fabius Tiro SW 3.n 149
Troubleshooting
When the battery is completely empty, Fabius
switches off and generates an acoustic alarm
signal (continuous tone for approx. 30 seconds).
All customized settings, including the alarm limits,
that deviate from the default settings are lost.
The following ventilation modes are still possible:
Manual ventilation
Spontaneous breathing
All pneumatic functions of Fabius are still
available:
APL valve
Pressure gauge for the airway pressure
Pressure gauge for gas cylinders and central
gas supply
Fresh-gas supply and anesthetic agent delivery
–S-ORC
Flow control valves for O
2, Air, and N2O
WARNING
Incorrect patient settings
When the power supply is restored and
Fabius is restarted, all ventilation and alarm
settings are reset to default settings.
After the restart of Fabius, check all settings
and adjust to the patient if necessary.