Instructions for use Fabius Tiro SW 3.n 207
Technical data
Dräger recommends complying with IEC 80001-1
(risk management for IT networks with medical
devices).
Serial interfaces
The following interfaces are supported:
RS232 interfaces complying with EIA RS--232
(CCITT V.24/V.28) for the following
applications:
MEDIBUS, MEDIBUS.X
Connections to medical devices from other
manufacturers
Electrical requirements of connected devices
and networks
The serial port is only appropriate for connecting
devices or networks that have a nominal voltage
on the network side of max. 24 V DC and meet the
requirements of one of the following standards:
IEC 60950-1: Ungrounded SELV circuits
IEC 60601-1 (2nd edition or later): Exposed
secondary electrical circuits