Annex
212 Instructions for use Fabius Tiro SW 3.n
Close the Air flow control valve. Open the
N
2O flow control valve completely. Check
whether the electronic N2O fresh-gas flow
display shows at least 10 L/min.
Check whether the float of the total flow
tube moves upward.
Shut off O
2 supply:
– Remove the O
2 connection of the cen-
tral gas supply.
– Close the O
2 cylinder valve.
– Check whether the red LED for low O
2
supply pressure is flashing.
–N
2O flow is interrupted.
Check whether the float of the total flow
tube shows 0 L/min.
Restore O
2 supply:
–N
2O flow is present.
Set the O
2 flow control valve to 1.5 L/min.
N
2O gas delivery = 3 L/min to 5 L/min.
Close the O
2 flow control valve:
–N
2O flow is interrupted.
Open the Air flow control valve. Check
whether the electronic Air fresh-gas flow
display shows at least 10 L/min.
Close all flow control valves.
Calibration of the sensors
Remove the O
2 sensor housing from the
cover of the inspiratory valve.
Expose the O
2 sensor to ambient air for
2 minutes.
Start the calibration.
Reinsert the O
2 sensor housing to the
cover of the inspiratory valve.
Calibrate the flow sensor
Check type of gas
Set the O
2 flow control valve to 3 L/min.
Check whether the measured O2 concen-
tration is at approx. 90 to 100 Vol%.
Close the O
2 flow control valve.
Vapor 19.n, Vapor 2000/3000 (Tec 5)
P Vaporizer is securely connected, locked,
and is hanging vertically.
P Control dial is in position 0.
P Filling level is between the minimum mark
and the maximum mark.
P Safety filling device:
– Sealing block is inserted and closed
tightly.
– Filling inlet is locked.
– Drainage valve closed.
P Filling device Quik Fil:
– Drainage valve closed.
– Sealing cap is tightly closed.
P Dräger Fill filling system:
– Drainage valve closed.
– Sealing cap is tightly closed.
D-Vapor (optional)
P Vaporizer is securely connected, locked
and is hanging vertically.
P Control dial is in position 0.
P Filling level is between the minimum mark
and maximum mark.
P Operation LED is lit.
Anesthetic vaporizers with Selectatec connec-
tion
P Vaporizer is securely connected, locked
and is hanging vertically.
P Control dial is in position 0.
P Filling level is between the minimum mark
and maximum mark.