Assembly and preparation
64 Instructions for use Fabius Tiro SW 3.n
Filling and mounting
1 After any cleaning, push the absorber insert
back into the absorber container (D)
completely.
2 Fill CO
2 absorber with fresh soda lime to the
upper mark.
Dräger recommends the use of
Drägersorb 800 Plus or Drägersorb FREE.
If conventional, non-pelletized soda lime is used,
then a soda lime dust filter
*
must be used.
3 Insert a new soda lime dust filter (E). Only use
soda lime dust filters indicated in the list of
accessories.
Only use undamaged filters, as exterior
damage to the filter decreases protection!
4 Place the CO
2 absorber (F) below the compact
breathing system in position and turn
counterclockwise to the final position stop.
21021
WARNING
Risk of hypoventilation
Reuse of the soda lime dust filter can
increase filter resistance and impair the
ventilation function of Fabius.
Replace the soda lime dust filter each time
the soda lime is replaced.
NOTE
Make sure that no soda lime residues are
between the seals and the seal surfaces. This
type of residue can cause system leakages.
D
*optional
20977
CAUTION
Risk of patient recovering consciousness
If the CO
2 absorber is not correctly locked into
place, system leakage may occur.
Make sure that the absorber container is
connected tightly to the compact breathing
system during the ventilation.
E