Maintenance Intervals
The workstation and its components must be cleaned and disinfected before every maintenance operation -
even when returning for repair!
Water separator Replace when dirty or when
the message CO2 line?! is displayed
(if the sample line is correctly installed and free of any blockage).
Can be disposed of as domestic waste.
O2 sensor Replace when calibration is no longer possible or when
the message FIO2 INOP! is displayed.
Disposal, see page 118.
Flow sensor Replace when calibration is no longer possible or when
the message FLOW INOP! is displayed.
Can be incinerated with little pollution at temperatures of more
than 800 °C.
Bacterial filter for secretion aspiration Replace after two weeks.
Must be disposed of as infectious special waste.
Can be incinerated with little pollution at temperatures of more
than 800 °C.
Bacterial filter Replace every six months.
for measured gas recirculation Can be incinerated with little pollution at temperatures of more
than 800 °C.
Cooling air filter (set of 3) Should be cleaned and thoroughly dried or replaced every month.
Must be replaced after one year at the latest.
Can be disposed of as domestic waste.
Lithium battery for data backup (set of 2) Must be replaced by professionals after two years.
Must be disposed of in accordance with local waste disposal regulations.
Lead gel battery in power pack (set of 2) Julian should be used in battery mode without a patient at least every
4 weeks.
A fully charged battery should keep the workstation operating for at least
30 minutes.
If not, get lead gel battery (set of 2) replaced by professionals.
Must be replaced by professionals after 2 years.
Must be disposed of in accordance with local waste disposal regulations.
Optical measuring set for determining
the anaesthetic gas concentration Must be checked by professionals every six months.
Time-keeper RAM Must be replaced by professionals after three years.
Must be disposed of in accordance with the local waste disposal
regulations.
Inspection and maintenance Must be carried out by professionals every six months
Maintenance Intervals
116