Cleaning the cooling air filters
clean filters monthly.
1 Remove all three cooling air filters from their mount.
Clean in warm water to which a detergent has been
added; dry thoroughly.
Insert the cooling air filters in their mount without
creasing.
Replace the cooling air filters after not more than one
year. The filters can be disposed of as domestic
waste.
Replacing the water separator
When soiled
or
when the message
CO2 line? ! is displayed
(if the sample line is correctly installed and free of any
blockage).
Grip the water separator by its sides and pull it out.
Push the new water separator into the holder as far as
possible.
The old water separator can be disposed of as
domestic waste.
Replacing the "Waterlock" water trap:
See separate Instructions for Use.
Replacing the O2 sensor
When the message
FIO2 INOP! is displayed
or
when the sensor can no longer be calibrated.
1 Remove the screw.
2 Remove the spent O2 sensor from the screw and
insert a new O2 sensor in the screw.
1 Refit the screw.
Dispose of the spent O2 sensor, see page 118.
117
Maintenance Intervals
Cleaning the cooling air filters
Replacing the water separator
Replacing the O2 sensor
1
2
1
1
1