1 Press the »MAN/SPONT« key,
2 and confirm with the rotary knob.
Display (example):
O2 flush
For flushing and rapidly filling the breathing system
and breathing bag with O2 while bypassing the
Vapor unit.
3 Press the »O2 +« button.
O2 flows into the breathing system without
anaesthetic gas as long as the button is held down.
Operation
Selecting the ventilation mode