Operation
Selecting the ventilation mode
MAN/SPONT
Certain alarms are automatically switched off in
MAN/SPONT ventilation mode in order to avoid artefacts.
The deactivated alarms are identified by the grey back-
ground in the table.
– – : The default setting on delivery is outside the
monitoring range and the corresponding alarm limit
is switched off.
To set the values of the alarm limits during operation see
page 60.
To set the default alarm limits see page 86.
Alarm limits in MAN/SPONT
mode
Default setting
on delivery
SpO2 ON
ON
– –
92
Pulse
H
ON
ON
120
50
etCO2 ON
ON
50 (mmHg)
– –
FiCO2 OFF
MV OFF
OFF
FiO2 OFF
ON
20
Fi Hal. ON
OFF
1.5
– –
Fi Iso. ON
OFF
2.3
– –
Fi Enf. ON
OFF
3.4
– –
Fi Des. ON
OFF
12.0
– –
Fi Sev. ON
OFF
3.4
– –
PAW ON
OFF
40
– –
Apnoea pressure OFF
Apnoea flow OFF
Apnoea CO2
ON
activated after
60 seconds
_
_
_
_
_
_
_
_
_
_
_
>