F9, F9 Express Fetal & Maternal Monitor User Manual Fetal Monitoring Display
- 107 -
Figure 8-6 File Management
4 If the required record is not on the current page, select Next to view more records.
5 Move the cursor to select the required item, and then select Export in the pop-up item, and the
monitor exports this record to the USB disk. Or you can select Export All to export all the
records to the USB disk.
NOTE:
1 When the monitor is in the process of printing or is connected to the network, the files
cannot be exported.
2 To avoid influence caused to the real-time monitoring, the manufacturer advises
against plugging in the USB disk and exporting the data in the process of monitoring.
3 The USB disk is not a tool for long-term data storage. You should save the exported
files in a PC in time and clear the USB disk termly.
4 The monitor only supports those USB disks with FAT or FAT32 (recommended)
format, and with capacity not larger than 8G. You are advised to use the USB disk
provided by the manufacturer.
In the FetusData folder of the USB disk, a sub-folder named after the export date and time is
created when the export is performed. The exported records are saved in this sub-folder as .trc
files, named after the monitor started date, time and ID, e.g. 20100120-124936-12345.trc.
8.2.3.2 Removing USB Disk
After the export finishes, select Remove Disk on the File Management interface. Do not unplug
the USB disk until a message “The USB can now be safely removed.” is prompted.
If the message Failure is prompted, you should perform the above procedures again.
NOTE:
1 Make sure you perform the Remove Disk procedure, otherwise data lose or USB disk