Refer to Connect the parallel control and monitoring on page 27 for more detailed
instrucons of how to use the parallel interface.
A system operang with only a parallel connecon is not capable of adjusng the conguraon
sengs stored in the motor controller (for example, power limit seng or controlled venng
opons). In this case, all these features would be at their factory default sengs. It is possible
to manually adjust the standby speed if standby mode is selected, however, the motor
controller should be congured separately before ng the pump to the system. This is
covered in more detail in Parallel control with serial monitoring or serial conguraon on page
17 and Motor controller conguraon (serial conguraon) on page 18.
Serial control and monitoring
The serial communicaons link provides complete control and monitoring by using just three
signal lines.
The serial data lines share the same connector pins as the parallel signals standby and fail. The
serial data lines can be
congured to provide an RS485 compliant or RS232 compable
interface by seng the posion of the RS485/RS232 slide switch (refer to Connecon for serial
control and monitoring on page 28).
The serial enable signal MUST be linked to 0 V for the system to accept commands from the
serial link. This is a safety feature and acts as an interlock. For pure serial control, the parallel
start signal will be le unconnected.
The motor controller will sll provide the normal and analogue signals on the logic interface
connector even when operang under serial control. The status of the normal signal and the
value of the system parameter on the analogue output can also be obtained by interrogang
the system status via the serial link.
For more informaon about the serial interface, refer to Connecon for serial control and
monitoring on page 28.
Serial control with parallel monitoring
Normal and analogue signals remain available when using serial control. It is possible to control
the pump via the serial link whilst monitoring these signals in the parallel interface.
The serial link uses the same connector pins as the parallel signals standby and fail so these
parallel control and monitoring signals are not available. The serial enable signal must be linked
to 0 V and the parallel start signal will be le unconnected.
Parallel control with serial monitoring or serial conguraon
Use this conguraon to operate the pump in parallel control mode, with the opon to adjust
the
conguraon sengs stored in the motor controller or to monitor operaonal status of the
pump through the serial link or via the USB service port.
If using the USB service port, in
conjuncon with Edwards nST PC soware, the serial enable
signal does not need to be linked to 0 V for the serial communicaons to take place. Whilst
operang under parallel control with the USB service port, all of the parallel control and
monitoring signals are available (as described in Parallel control and monitoring on page 16),
including the standby control line and fail monitoring line.
If using the serial link, the serial enable signal must be linked to 0 V for serial communicaons
to take place. Whilst operang under parallel control with the serial link acve, the parallel
start control signal is available (as described in Parallel control and monitoring on page 16) but
the standby control line is not since it is used as a serial data line.
If the serial enable line is deacvated at any me whilst the RS485/RS232 slide switch is in the
RS232 posion, the serial link should also be disconnected. Edwards suggests making a special
B8G0-00-880A - Overview of the nEXT85
Page 17