Error ash
posion
Error ash
code
Comments Acons
2 sLssss Controller failed internal
conguraon and calibraon
operaon.
Cycle the power to the pump
and see whether the error
code appears again. If it does,
contact the supplier or
Edwards.
3 ssLsss Failure to reach or maintain
half full speed within the
mer seng value.
Check whether the pump is
too hot or whether the inlet
pressure is too high.
4 sssLss Overspeed or overcurrent trip
acvated, or other hardware
fault.
Cycle the power to the pump
and see whether the error
code appears again. If it does,
contact the supplier or
Edwards.
5 ssssLs Pump internal measurement
system disconnected or
damaged.
Cycle the power to the pump
and see whether the error
code appears again. If it does,
contact the supplier or
Edwards.
6 sssssL Serial enable becomes
inacve following a serial start
command.
Re-acvate serial enable and
send a serial stop command to
clear the error code.
Note:
The alarm LED error ash sequence is capable of signaling mulple fail condions. For
example, error ash code sLssLs signies both error 2 (controller failed internal conguraon
and calibraon operaon) and error 5 (pump internal measurement system disconnected or
damaged).
Decoding system status words
If using the serial
communicaons link, addional informaon, that may be useful for fault
nding, can be accessed.
When sending a query to monitor measured motor speed, the pump also returns a system
status word.
The send command is as follows:
Command
? V
8 5 2
cr
The reply will be as follows, where the rst part returned is a number refers to motor
rotaonal speed in revoluons per second (Hz) and the second part is the system status word:
Reply
= V
8 5 2
sp
d d d d
;
h h h h h h h h
cr
The system status word returned is made up of 8 hexadecimal digits. To decode this word,
convert each digit into a 4-digit binary number. Follow the example below:
2
2 8 3 0 0 2 2
B8G0-00-880A - Troubleshoong
Page 70