Pump run me
The run hours and recommended service me of the rotor in the nEXT85 pump can be
monitored.
Send the following query:
Command ? V
8 8 3
cr
The reply will be as follows, where the rst number is the hours run by the rotor and the
second is the number of hours unl service is recommended:
Reply = V
8 8 3
sp
d d d d d d
;
d d d d d d
cr
Pump cycles
The number of start-stop cycles completed and the number remaining
unl the next service is
due can be monitored.
Send the following query:
Command
? V
8 8 4
cr
The reply will be as follows, where the rst number is the start-stop cycles completed by the
pump and the second is the number of start-stop cycles unl service is recommended:
Reply
= V
8 8 4
sp
d d d d d d
;
d d d d d d
cr
Bearing run me
The run hours and recommended service me of the bearing in the nEXT85 pump can be
monitored.
Send the following query:
Command
? V
8 8 5
cr
The reply will be as follows, where the rst number is the hours run by the bearing and the
second is the number of hours unl service is recommended:
Reply
= V
8 8 5
sp
d d d d d d
;
d d d d d d
cr
Oil cartridge run me
The run hours and recommended service me of the oil cartridge in the nEXT85 pump can be
monitored.
Send the following query:
Command
? V
8 8 6
cr
The reply will be as follows, where the rst number is the hours run by the oil cartridge and the
second is the number of hours unl service is recommended:
B8G0-00-880A - Maintain the nEXT85
Page 61