cable for serial communicaons that includes a link between serial enable and 0 V. This way,
serial enable is automacally acvated when the cable is connected and then deacvated
when the cable is removed.
Motor controller conguraon (serial conguraon)
All the conguraon sengs stored within the motor controller are retained when power to
the pump is removed so that it is possible to use a separate system to congure the motor
controller before
ng the pump to the applicaon.
This gives the benet of tailoring the pump funconality to a customer applicaon and allows
the pump to be operated using a simple parallel interface system.
To congure the pump, either use a customer simple serial system, via the serial link, or use
the Edwards TIC Turbo Controller, TIC Turbo Instrument Controller or TAG Controller. The TICs
have a feature which allows storage of a nEXT85 pump’s conguraon. The conguraon can
then be downloaded to another nEXT85 pump. This is useful when conguring a number of
nEXT85 pumps with the same sengs before they are ed to a system.
The USB service port can also be used in conjuncon with the Edwards nST PC soware to
congure the nEXT85 pump. The conguraon can then be downloaded to another nEXT85
pump, in the same way it can with a TIC, which is useful when
conguring a number of nEXT85
pumps.
B8G0-00-880A - Overview of the nEXT85