If the nEXT85 pump receives characters that are not framed inside start and stop characters, it
will ignore them. Messages with the stop character missing will be discarded with no response
when a new start character is received. If the nEXT85 pump receives an unrecognisable
message between the start and stop characters, it will return an appropriate error message.
Refer to Mul-drop operaon on page 99 for more informaon about operang the nEXT85
pumps in mul-drop mode.
Command set
Table 45
Summary of commands that can be sent to the pump
Object name Command
Parameter
range
Factory
seng
Data
type
Units Comments
Node !S850 0..99 - decimal address Mul-drop address
?850 0 = disable mul-drop
address
99 = wild card
Pump type ?S851 8; - string chars Pump type
10; string chars DSP soware version
number
4 string chars (D39659610)
Full speed RPS (1500 for
nEXT85)
Pump control !C852 0 - decimal - Stop the pump
1 Start the pump
?V852 0..1800; - decimal RPS Measured motor speed
32-bits hex ags System status word
Time seng !S854 1..30 8 decimal minutes Timeout period for both
inial ramp
?S854 up and if speed drops
below 50%
Power limit
seng
!S855 50..120 80 decimal Was Link power maximum
?S855
Normal speed
seng
!S856 50..100 80 decimal % Normal speed as a
percentage of full speed
?S856
Standby speed
seng
!S857 55..100 70 decimal % Standby speed as a
percentage of full speed
?S857
Temperature
readings
?V859 0..100; - decimal °C Measured motor
temperature
0..100 decimal °C Measured motor
controller temperature
B8G0-00-880A - Technical Reference
Page 95