Page #96 background imageLoading...
Page #96 background image
Object name Command
Parameter
range
Factory
seng
Data
type
Units Comments
Link
parameter
readings
?V860 0..500; - decimal 0.1
Volts
Measured link voltage
0..300; decimal 0.1
Amps
Measured link current
0..15000 decimal 0.1
Was
Measured link power
Pump run
hours
?V862 0..65535 - decimal hours Hours run by pump
Temperature
readings 2
?V865 -200..200; - decimal °C Measured motor
temperature
-200..200; decimal °C Measured motor
controller temperature
-200..200 decimal °C Measured rotor
temperature
Factory
sengs
!S867 1 - - - Reset all conguraon
opons and parameters
to the factory sengs
PIC soware
version
?S868 10 - string chars Boot loader soware
version number
(D396596XXX)
Speed control !C869 0 - decimal - Set target speed to full
speed
1 Set target speed to
standby speed
Timer opons !S870 0 1 decimal - Timer = disabled
?S870 1
Timer = enabled
Note that the mer is
permanently enabled
on ramp-up.
Analogue
signal opons
!S871 0 0 decimal - Analogue output =
measured speed
?S871 1 Analogue output =
measured power
2 Analogue output =
measured motor temp.
3 Analogue output =
measured control temp.
4 Analogue output =
measured rotor temp.
B8G0-00-880A - Technical Reference
Page 96

Table of Contents

Other manuals for Edwards nEXT85

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards nEXT85 and is the answer not in the manual?

Edwards nEXT85 Specifications

General IconGeneral
BrandEdwards
ModelnEXT85
CategoryWater Pump
LanguageEnglish

Related product manuals