Bit number Status of bit
(in example)
We can deduce
13 0 Overspeed and overcurrent trip not acvated
14 0 Pump internal temperature measurement system is ne
15 0 Serial enable has not become inacve during serial control
Service informaon
When using the serial communicaons link, addional informaon about the pump, such as
pump type and internal controller soware versions, can be accessed. This informaon is
parcularly useful for service personnel to determine the model of the pump.
Send the following query to nd out pump type:
Command
? S
8 5 1
cr
The reply will be as follows, where string 1 is the pump type, string 2 is the DSP soware
version number and string 3 is the designated full speed of the pump (in revoluons per
second):
Reply
= S
8 5 1
sp
String
1
;
String
2
;
String
3
cr
Send the following query to nd out the boot loader soware version:
Command
? S
8 6 8
cr
The reply will be as follows, where string1 is the boot loader soware version number:
Reply
= S
8 6 8
sp
String
1
cr
B8G0-00-880A - Troubleshoong
Page 73