Connect the serial interface to the customer control equipment
The serial interface is available in RS485 or RS232 opons which can be selected using the slide
switch located adjacent to the main power lead.
Refer to Figure 4 on page 30. To adjust the slide switch, toggle the slide switch using a small
tool. The motor controller default seng is RS232 serial interface. Be sure to replace the round
seal in order to ensure the IP rang of the pump is sll achieved.
CAUTION:
When connecng the pump to a PC, note that the 0 V pin on the RS232 connector may be
connected to earth through the PC. If this is the case, ensure that the 0 V rail of the 24 - 48
V d.c. supply is not also connected to earth at some other point such as at the power
supply. If the 0 V rail of the 24 - 48 V d.c. supply will not be connected to earth at the PC,
an opto-isolated interface to the PC should be used.
The nEXT85 pump can connect directly to the RS485 or RS232 serial input on the control
equipment or a PC as shown in Figure 5 on page 30 and Figure 6 on page 31. In this
conguraon the PC is the serial link master and the nEXT85 pump is the slave. The distance
over which the serial link will work is dependent on any dierence in voltage between the 0 V
at the sending and receiving end. If the 0 V reference at the receiving end is within 0.3 V of the
0 V Control Reference pin on the nEXT85 pump control connector then the serial link should be
capable of operang at distances up to 6 m. An interface circuit external to the nEXT85 pump
may be required for longer distances.
The soware in the nEXT85 pump is capable of operang with several pumps connected to a
single serial link master. This is referred to as mul-drop mode. The RS485 opon is
recommended for mul-drop mode. With the RS232 opon selected, some addional
hardware will be required to link several nEXT85 pump units to a single serial link master. A
concept drawing of one possible arrangement is shown in Figure 18 on page 100. Conversely
when the RS485 opon is selected, connecng several nEXT85 pumps to a single master
becomes a simple wiring exercise as shown in Figure 19 on page 101.
B8G0-00-880A - Install the nEXT85
Page 29

Table of Contents