Mixed parallel and serial operaon
In mixed parallel and serial operaon the pump may receive commands from both serial and
parallel interfaces.
To understand how these commands control the pump, refer to Figure 7 on page 32. The pump
will power up with ‘None in Control’. From this state a parallel start signal or a serial start
command may be received,
resulng in the pump moving to parallel control mode or serial
control mode respecvely. Serial start commands will only be received if the serial enable line
is acve.
The state of the serial enable line may be switched between acve and inacve whilst in mixed
parallel and serial operaon. The primary funcon of the serial enable line is to enable the
serial link. It has no direct eect on the control mode. The pump will receive and respond to
serial commands whenever the serial enable line is acve, and conversely will not receive or
respond to serial commands when the serial enable line is inacve.
Availability of the parallel standby and fail signals depends on the state of the serial enable line
and the posion of the RS485/RS232 switch as described in Table 9 on page 52.
Table 9
Serial enable matrix
Switch Serial enable acve Serial enable inacve
RS232
Standby and fail lines are used for
RS232 data.
Standby and fail lines are used for
parallel Standby control and fail
indicaon.
Parallel standby control and fail
indicaon are disabled.
Serial communicaons are disabled.
RS485
Standby and fail lines are used for
RS485 data.
Standby and fail lines are in (RS485)
high impedance state.
*
Parallel standby control and fail
indicaon are disabled.
Serial communicaons are disabled.
*
Pump target speed will switch between run speed and standby speed in response to any data driven onto the
RS485 bus by other devices connected to the bus. Acvaon of the parallel start line not recommended with the
RS485/232 switch in RS485 posion and serial enable Inacve. Deacvaon of the serial enable line is not
recommended with the pump running following a start from the parallel start line when the RS485/232 switch is in
the RS485
posion.
In parallel control mode, the pump will not accept serial stop commands but will accept all
other serial commands. When serial enable is acve, the pump will run at standby speed if it
has been commanded to do so by a serial standby command. When Serial enable is inacve
and the RS485/RS232 switch is in RS232 posion, the pump will run at standby speed when the
parallel standby line is acve.
In serial control mode, the state of the parallel Start line will be ignored but the serial enable
line provides an interlock funcon as shown in Figure 9 on page 53. This interlock funcon
only operates with serial start commands and therefore only in serial control mode. In serial
control mode, the pump cannot be commanded to standby speed by the standby line, instead
a serial standby command must be used.
The parallel normal and analogue output signals provide valid pump status informaon at all
mes under mixed parallel and serial operaon. The analogue output voltages are described in
Table 43 on page 91.
B8G0-00-880A -
Operaon of the nEXT85