Command ? S
8 5 0
cr
If the reply is as shown below, the pump has mul-drop mode disabled:
Reply = S
8 5 0
sp
0
cr
If the pump already has a mul-drop address, there will be no reply. Communicate with the
pump in mul-drop message protocol. Refer to Mul-drop operaon on page 99 for more
informaon about mul-drop mode and mul-drop message protocol.
Use the following query (using wild card address 99 which means ‘any’ node) to nd out the
mul-drop address of the nEXT85 pump:
Command
#
9 9
:
9 9
? 5
8 5 0
cr
The reply will be as follows, where dd denotes the mul-drop address of the pump:
Reply
#
9 9
:
9 9
= S
8 5 0
sp
d d
cr
Mul-drop mode can be disabled by assigning the pump an address 0. To do this, send the
following command (where dd denotes the mul-drop address of the pump and XX denotes
the address of the node that is sending the command):
Command
#
d d
:
x x
1 S
8 5 0
sp
0
cr
The reply will be in the following format:
Reply
#
x x
:
d d
* S
8 5 0
sp
0
cr
Once mul-drop mode is disabled, the pump will no longer respond to mul-drop commands.
Congure the pump using a TIC
The nEXT85 pump can be congured using the Edwards TIC Turbo and Instrument Controller or
TIC Turbo Controller.
Refer to TIC or TAG logic interface connecons on page 26.
It is possible to set the following parameters of the nEXT85 pump using the TIC:
• Power limit seng
• Vent opons and vent type, including running a fan from the motor controller
• Standby speed seng
• Normal speed seng
• Timer sengs - both enabling/disabling the mer AND seng the meout period
• Electronic braking opons
• Factory default sengs
For informaon on how to perform these sengs, refer to the TIC Turbo and Instrument
controller or TIC Turbo Controller Instrucon Manuals.
Note:
It is not possible to congure the analogue output opons using the TIC. There is also no
facility within the TIC to assign a mul-drop address to the nEXT85 pump.
B8G0-00-880A - Conguraon