The permanently enabled fan seng is now stored in memory within the nEXT85 pump.
When the pump is shipped, it is set up to run a vent valve. Send a query to nd out what the
seng is as follows:
Command ? S
8 6 4
cr
The reply will be in the following format:
Reply * S
8 6 4
sp
d
cr
If the character 'd' is 8, then the fan is enabled. If it is anything else, congure the motor
controller to run the fan.
Controlled venng opons
If the motor controller is used to automacally control a vent valve, there are a number of
venng opons available.
Refer to Table 6 on page 46 for available venng opons and refer to Table 7 on page 47 for
available valve types.
To set a vent opon on auxiliary output 1, send the following command, (where the character
'd' refers to the opon number shown in Table 6 on page 46):
Command
! S
8 5 3
sp
d
cr
The reply will be in the following format:
Reply
* S
8 5 3
sp
r
cr
The venng opon is now stored in memory within the nEXT85 pump.
To check the venng opon that is set, send a query as follows:
Command
? S
8 5 3
cr
The reply will be in the following format:
Reply
= S
8 5 3
sp
d
cr
To set the valve type on auxiliary output 1, send the following command, (where the character
'd' refers to the opon number shown in Table 7 on page 47):
Command
! S
8 7 7
sp
d
cr
The reply will be in the following format:
Reply
* S
8 7 7
sp
r
cr
The valve type is now stored in memory within the nEXT85 pump, but a power cycle is required
before the auxiliary 1 output circuitry adjusts to the stored request.
To check the venng opon that is set, send a query as follows:
Command
? S
8 7 7
cr
B8G0-00-880A - Conguraon
Page 38