Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block
Endress+Hauser 109
3 Transducer Block
The Transducer Blocks contain all the measurement- and device-specific parameters of the
flowmeter. All the settings directly connected with the flow measurement/application are made
here. They form the interface between the sensor-specific measured value preprocessing and the
Analog Input function blocks required for automation.
A Transducer Block allows you to influence the input and output values of a function block. The
parameters of a Transducer Block include information on the sensor type, sensor configuration,
physical units, calibration, damping, diagnosis, etc. as well as the device-specific parameters. The
device-specific parameters and functions are split into several Transducer Blocks, each covering
different task areas.
"Flow" Transducer Block/base index 1400:
This block contains all the flow-specific parameters and functions, e.g. calibration functions, sensor
data, etc. → Page 113
"Diagnosis" Transducer Block/base index 1600:
This block contains all the parameters for system diagnosis, e.g. current system status etc.
→ Page 113
"Display" Transducer Block/base index 1800:
This block contains all the parameters for configuring the local display
→ Page 134
"Totalizer" Transducer Block/base index 1900:
This block contains all the parameters for configuring the totalizers
→ Page 151
"Calculated Density" Transducer Block/base index 2000:
This block contains all the parameters for configuring density measurement
→ Page 151
"Viscosity" Transducer Block/base index 2100:
This block contains all the parameters for configuring viscosity measurement
→ Page 151
"Advanced Diagnostics" Transducer Block/base index 2200:
This block contains all the parameters for configuring advanced diagnostics
→ Page 151