EasyManua.ls Logo

Endress+Hauser Proline Promass 83 - Function Group REFERENCE PARAMETER

Endress+Hauser Proline Promass 83
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 7 Block BASIC FUNCTION
Endress+Hauser 55
7.2.3 Function group REFERENCE PARAMETER
EPD EXCITATION
CURRENT MAXIMUM
(6426)
Use this function to activate the empty pipe detection (EPD).
In the event of inhomogeneous fluids or air bubbles, the excitation current of the
measuring pipes increases. If the excitation current set in this function is exceeded,
error message #700 is output similar to the functions "EPD VALUE LOW (6423)" and
"EPD VALUE HIGH (6424)".
User input:
5-digit floating-point number
Factory setting:
100.00 mA (deactivated)
! Note!
The function is not activated until a value under 100 mA is input. Entering the value
100 mA deactivates the function.
BASIC FUNCTION G
FOUNDATION FIELDBUS GGA
PROCESS PARAMETER GIA
CONFIGURATION 640
EPD PARAMETER 642
REFERENCE PARAMETER 646
Function description
BASIC FUNCTION PROCESS PARAMETER REFERENCE PARAMETER
CORRECTED VOLUME
CALCULATION
(6460)
This function is used to set the reference density for calculating the corrected volume
flow.
Options:
FIXED REFERENCE DENSITY
CALCULATED REFERENCE DENSITY
Factory setting:
CALCULATED REFERENCE DENSITY
FIXED REFERENCE
DENSITY
(6461)
! Note!
This function is not available unless the FIXED REFERENCE DENSITY setting was
selected in the CORRECTED VOLUME CALCULATION function (6460).
In this function, a fixed value for the reference density can be entered, with which the
corrected volume flow or corrected volume is calculated.
User input:
5-digit floating-point number
Factory setting:
1.0000 kg/Nl
Function description
BASIC FUNCTION PROCESS PARAMETER EPD PARAMETER

Table of Contents

Other manuals for Endress+Hauser Proline Promass 83

Related product manuals