7 Block BASIC FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus
58 Endress+Hauser
DENSITY SETPOINT 2
(6485)
Use this function to enter the density setpoint value for the second fluid for which you
want to carry out field density adjustment.
User input:
5-digit floating-point number, incl. units
! Note!
• The preset density entered here should not vary from the actual fluid
density by a more than ±10%.
• The difference between the density setpoint values must be at least 0.2 kg/l.
• The appropriate unit is taken from the SYSTEM UNITS group
( → Page 14).
MEASURE FLUID 2
(6486)
In this function the current density of the second fluid is measured for density
adjustment.
Options:
CANCEL
START
DENSITY ADJUSTMENT
(6487)
With this function a density adjustment can be carried out on site. The density
adjustment values will thus be recalculated and stored. This ensures that the values
dependent on density calculations (e.g. volume flow) are as accurate as possible.
! Note!
Before carrying out a density adjustment, please refer to BA059D/06/en "Proline
Promass 83 FOUNDATION Fieldbus Operating Instructions" where a detailed
description of the density adjustment is given.
Options:
CANCEL
DENSITY ADJUST
Factory setting:
CANCEL
RESTORE ORIGINAL
(6488)
With this function the original density coefficients determined at the factory are
restored.
Options:
NO
YES
Factory setting:
NO
Function description
BASIC FUNCTION → PROCESS PARAMETER → ADJUSTMENT