8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus
82 Endress+Hauser
TUBE DAMPING
FLUCTUATION
(7491)
The measured fluctuation of the tube damping appears on the display.
Display:
5-digit floating-point number
MINIMUM
TUBE DAMPING
FLUCTUATION
(7492)
The lowest value of the tube damping since the saved values were last reset appears on
the display.
Display:
5-digit floating-point number
MAXIMUM
TUBE DAMPING
FLUCTUATION
(7493)
The highest value of the tube damping fluctuation since the saved values were last reset
appears on the display.
Display:
5-digit floating-point number
HISTORY TUBE
DAMPING
FLUCTUATION
(7494)
The last ten values of the tube damping fluctuation since the saved values were last reset
appear on the display.
Display:
5-digit floating-point number
DEVIATION
TUBE DAMPING
FLUCTUATION
(7495)
This function displays the deviation between the measured fluctuation of the tube
damping and the reference values (FACTORY or USER) selected in the function
SELECT REFERENCE CONDITION (7402).
Display:
5-digit floating-point number
WARNING LEVEL
(7496)
! Note!
This function is not available unless ON was selected in the WARNING MODE function
(7403).
Use this function to specify a limit value for the tube damping fluctuation. A notice
message is generated if the limit value is exceeded.
User input:
0 to 99999
Factory setting:
1000
Function description
SPECIAL FUNCTION ADVANCED DIAGNOSTICS TUBE DAMPING FLUCTUATION