7 Block BASIC FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus
52 Endress+Hauser
7.2 Group PROCESS PARAMETER
7.2.1 Function group CONFIGURATION
BASIC FUNCTION G
FOUNDATION FIELDBUS GGA
PROCESS PARAMETER GIA
CONFIGURATION 640
Function description
BASIC FUNCTION PROCESS PARAMETER CONFIGURATION
ASSIGN LOW FLOW
CUT OFF
(6400)
Use this function to assign the switching point for low flow cut off.
Options:
OFF
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW
Factory setting:
MASS FLOW
ON-VALUE LOW
FLOW CUT OFF
(6402)
Use this function to assign a value to the switch-on point for low flow cut off.
Low flow cut off is active if the setting is a value not equal to 0. The sign of the flow
value is highlighted on the display to indicate that low flow cut off is active.
User input:
5-digit floating-point number, (unit)
Factory setting:
Depends on nominal diameter and country Page 90 ff.
! Note!
The appropriate unit is taken from the function group SYSTEM UNITS (Page 14).
OFF-VALUE LOW
FLOW CUT OFF
(6403)
Use this function to enter the off-value (b) for low flow cut off.
Enter the off-value as a positive hysteresis (H) from the on-value (a).
User input:
Integer 0 to 100%
Factory setting:
50%
A0003882
m = On-value, n =Off-value
a Low flow cut off is switched on
b Low flow cut off is switched off (a + a
H)
H Hysteresis value: 0 to 100%
Low flow cut off active
Q
Flow
t
H
Q
m
n
b
a