Device Functions Proline Promass 83 FOUNDATION Fieldbus 8 Block SPECIAL FUNCTION
Endress+Hauser 77
8.2.6 Function group TEMPERATURE
SPECIAL FUNCTION H
⇒
DENSITY FUNCTIONS HAA
⇓
ADVANCED DIAGNOSTICS HEA
⇒
CONFIGURATION 740
⇓
ACQUISITION 741
⇓
MASS FLOW 742
⇓
DENSITY 743
⇓
REFERENCE DENSITY 744
⇓
TEMPERATURE 745
Function description
SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → TEMPERATURE
! Note!
The appropriate unit is taken from the UNIT TEMPERATURE function (0422).
REFERENCE VALUE
TEMPERATURE
(7450)
The reference value for the temperature appears on the display.
Display:
5-digit floating-point number, including unit and sign
TEMPERATURE
(7451)
The currently measured temperature appears on the display.
Display:
5-digit floating-point number, including unit and sign
MINIMUM
TEMPERATURE
(7452)
The lowest temperature value since the saved values were last reset appears on the
display.
Display:
5-digit floating-point number, including unit and sign
MAXIMUM
TEMPERATURE
(7453)
The highest temperature value since the saved values were last reset appears on the
display.
Display:
5-digit floating-point number, including unit and sign
HISTORY TEMPERATURE
(7454)
The last ten temperature values since the saved values were last reset appear on the
display.
Display:
5-digit floating-point number, including unit and sign
DEVIATION
TEMPERATURE
(7455)
This function displays the deviation between the currently measured temperature and the
reference values (FACTORY or USER) selected in the function SELECT REFERENCE
CONDITION (7402).
Display:
5-digit floating-point number, including unit and sign
WARNING LEVEL
(7456)
! Note!
This function is not available unless ON was selected in the WARNING MODE function
(7403).
Use this function to specify a limit value for the temperature. A notice message is
generated if the limit value is exceeded.
User input:
0 to 99999 °C
Factory setting:
100 °C