Device Functions Proline Promass 83 FOUNDATION Fieldbus 8 Block SPECIAL FUNCTION
Endress+Hauser 75
8.2.4 Function group DENSITY
SPECIAL FUNCTION H
DENSITY FUNCTIONS HAA
ADVANCED DIAGNOSTICS HEA
CONFIGURATION 740
ACQUISITION 741
MASS FLOW 742
DENSITY 743
Function description
SPECIAL FUNCTION ADVANCED DIAGNOSTICS DENSITY
! Note!
The appropriate unit is taken from the UNIT DENSITY function (0420).
REFERENCE VALUE
DENSITY
(7430)
The reference value for the density appears on the display.
Display:
5-digit floating-point number, incl. units
DENSITY
(7431)
The measured density appears on the display.
Display:
5-digit floating-point number, incl. units
MINIMUM DENSITY
(7432)
The lowest density value since the saved values were last reset appears on the display.
Display:
5-digit floating-point number, incl. units
MAXIMUM DENSITY
(7433)
The highest density value since the saved values were last reset appears on the display.
Display:
5-digit floating-point number, incl. units
HISTORY DENSITY
(7434)
The last ten density values since the saved values were last reset appear on the display.
Display:
5-digit floating-point number, incl. units
DEVIATION DENSITY
(7435)
This function displays the deviation between the measured density and the reference
values (FACTORY or USER) selected in the function SELECT REFERENCE CONDITION
(7402).
Display:
5-digit floating-point number, incl. units
WARNING LEVEL
(7436)
! Note!
This function is not available unless ON was selected in the WARNING MODE function
(7403).
Use this function to specify a limit value for the density. A notice message is generated if
the limit value is exceeded.
User input:
0 to 99999 [%]
Factory setting:
100%