8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus
80 Endress+Hauser
8.2.9 Function group OPERATING FREQUENCY FLUCTUATION
WARNING LEVEL
(7476)
! Note!
This function is not available unless ON was selected in the WARNING MODE function
(7403).
Use this function to specify a limit value for the electrodynamic sensors. A notice message
is generated if the limit value is exceeded.
User input:
0 to 99999%
Factory setting:
100%
SPECIAL FUNCTION H
DENSITY FUNCTIONS HAA
ADVANCED DIAGNOSTICS HEA
CONFIGURATION 740
ACQUISITION 741
MASS FLOW 742
DENSITY 743
REFERENCE DENSITY 744
TEMPERATURE 745
TUBE DAMPING 746
ELEC.-DYN. SENSORS 747
OPERATING FREQ.
FLUCT.
748
Function description
SPECIAL FUNCTION ADVANCED DIAGNOSTICS OPERATING FREQUENCY FLUCTUATION
REFERENCE VALUE
OPERATING
FREQUENCY
FLUCTUATION
(7480)
The reference value for the fluctuation of the operating frequency appears on the display.
Display:
5-digit floating-point number, Hz
OPERATING
FREQUENCY
FLUCTUATION
(7481)
The measured fluctuation of the operating frequency appears on the display.
Display:
5-digit floating-point number, Hz
MINIMUM
OPERATING
FREQUENCY
FLUCTUATION
(7482)
The lowest value of the operating frequency fluctuation since the saved values were last
reset appears on the display.
Display:
5-digit floating-point number, Hz
MAXIMUM
OPERATING
FREQUENCY
FLUCTUATION
(7483)
The highest value of the operating frequency fluctuation since the saved values were last
reset appears on the display.
Display:
5-digit floating-point number, Hz
Function description
SPECIAL FUNCTION ADVANCED DIAGNOSTICS ELECTRODYNAMIC SENSORS