Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Block MEASURED VARIABLES
Endress+Hauser 13
CORRECTED CARRIER
VOLUME FLOW
(0029)
! Note!
This function is not available unless % MASS / % VOLUME was selected in the
DENSITY FUNCTION function (7000).
Use this function to display the currently measured corrected volume flow of the carrier
fluid. Carrier fluid = transporting liquid (e.g. water).
Display:
5-digit floating-point number, including unit and sign
%-BLACK LIQUOR
(0030)
! Note!
This function is not available unless %-BLACK LIQUOR was selected in the
DENSITY FUNCTION function (7000).
The concentration in %-BLACK LIQUOR is displayed.
Display:
5-digit floating-point number, incl. units
° BAUME
(0031)
! Note!
This function is not available unless °BAUME was selected in the DENSITY FUNCTION
function (7000).
The concentration in °BAUME is displayed.
Display:
5-digit floating-point number, incl. units
° API
(0033)
! Note!
This function is not available unless °API was selected in the DENSITY FUNCTION
function (7000).
The concentration in °API is displayed.
Display:
5-digit floating-point number, incl. units
° PLATO
(0034)
! Note!
This function is not available unless °PLATO was selected in the DENSITY FUNCTION
function (7000).
The concentration in °PLATO is displayed.
Display:
5-digit floating-point number, incl. units
° BALLING
(0035)
! Note!
This function is not available unless °BALLING was selected in the DENSITY FUNCTION
function (7000).
The concentration in °BALLING is displayed.
Display:
5-digit floating-point number, incl. units
° BRIX
(0036)
! Note!
This function is not available unless °BRIX was selected in the DENSITY FUNCTION
function (7000).
The concentration in °BRIX is displayed.
Display:
5-digit floating-point number, incl. units
Function description
MEASURED VARIABLES → MEASURING VALUES → ADDITIONAL VALUES