8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus
72 Endress+Hauser
8.2 Group ADVANCED DIAGNOSTICS
8.2.1 Function group CONFIGURATION
SPECIAL FUNCTION H
DENSITY FUNCTIONS HAA
ADVANCED DIAGNOSTICS HEA
CONFIGURATION 740
Function description
SPECIAL FUNCTION ADVANCED DIAGNOSTICS CONFIGURATION
REFERENCE
CONDITION USER
(7401)
Use this function to start determining the user reference status. The following values are
determined:
MASS FLOW
•DENSITY
REFERENCE DENSITY
TEMPERATURE
TUBE DAMPING
ELECTRODYNAMIC SENSORS
OPERATING FREQUENCY FLUCTUATION
TUBE DAMPING FLUCTUATION
Options:
CANCEL
START
Factory setting:
CANCEL
SELECT REFERENCE
CONDITION
(7402)
Use this function to select the reference status which should be used to compare the
advanced diagnostics parameters (see function ACQUISITION MODE (7410)).
Options:
FACTORY
USER
Factory setting:
FACTORY
WARNING MODE
(7403)
Use this function to determine whether a warning should be generated when there
is a deviation between the reference status (FACTORY or USER, see function
SELECT REFERENCE CONDITION (7402)) and the current measuring values.
The values of the following functions are compared to the reference status :
MASS FLOW (7421)
DENSITY (7431)
REFERENCE DENSITY (7441)
TEMPERATURE (7451)
TUBE DAMPING (7461)
ELECTRODYNAMIC SENSORS (7471)
OPERATING FREQUENCY FLUCTUATION (7481)
TUBE DAMPING FLUCTUATION (7491)
Options:
OFF
ON
Factory setting:
OFF