8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus
76 Endress+Hauser
8.2.5 Function group REFERENCE DENSITY
SPECIAL FUNCTION H
DENSITY FUNCTIONS HAA
ADVANCED DIAGNOSTICS HEA
CONFIGURATION 740
ACQUISITION 741
MASS FLOW 742
DENSITY 743
REFERENCE DENSITY 744
Function description
SPECIAL FUNCTION ADVANCED DIAGNOSTICS REFERENCE DENSITY
! Note!
The appropriate unit is taken from the function UNIT REFERENCE DENSITY (0421) ( Page 17).
REFERENCE VALUE
REFERENCE DENSITY
(7440)
The reference value for the reference density appears on the display.
Display:
5-digit floating-point number, incl. units
REFERENCE DENSITY
(7441)
The measured reference density appears on the display.
Display:
5-digit floating-point number, incl. units
MINIMUM
REFERENCE DENSITY
(7442)
The lowest reference density value since the saved values were last reset appears on the
display.
Display:
5-digit floating-point number, incl. units
MAXIMUM
REFERENCE DENSITY
(7443)
The highest reference density value since the saved values were last reset appears on the
display.
Display:
5-digit floating-point number, incl. units
HISTORY
REFERENCE DENSITY
(7444)
The last ten reference density values since the saved values were last reset appear on the
display.
Display:
5-digit floating-point number, incl. units
DEVIATION
REFERENCE DENSITY
(7445)
This function displays the deviation between the measured reference density and the
reference values (FACTORY or USER) selected in the function SELECT REFERENCE
CONDITION (7402).
Display:
5-digit floating-point number, incl. units
WARNING LEVEL
(7446)
! Note!
This function is not available unless ON was selected in the WARNING MODE function
(7403).
Use this function to specify a limit value for the reference density. A notice message is
generated if the limit value is exceeded.
User input:
0 to 99999 [%]
Factory setting:
100%