Chapter 6: Preventive maintenance and testing
152 EST3X Technical Reference Manual
Required tools
• Slotted screwdriver, insulated
• Digital multimeter
• 1.1 kΩ,1 W resistor
• 12-inch (30.5 cm) jumper lead with alligator clips
• Commercial grade (1200 to 1500 W) hair blower (for testing heat detectors)
• Conventional vacuum cleaner (for cleaning SIGA and SIGA2 detectors)
• Signature Series Tool Kit (P/N SIGA-ST) (for cleaning SIGA detectors)
• Control panel door key
• System passwords (if any)
Signature device maintenance tips
Detectors
When removing one detector at a time, wait 1-minute after replacing the first
detector before removing the next detector. This gives the system time to
recognize and remap the first detector before generating a trouble condition
caused by removing the second detector.
CO maintenance alert
In addition to displaying a maintenance alert when the photo element dirtiness is
at or above 80%, the loop controller displays a maintenance alert when the CO
sensor module is at or below 6 months until end-of-life. If both elements are at or
above these thresholds, there is only one maintenance alert.
Once the dirtiness threshold is at 100%, a dirty detector trouble displays for the
photo element. Once there are zero months until end-of-life, the panel displays
the CO end-of-life trouble message.
CO maintenance report
The CO sensor module has a life span of 6 years. After 6 years, the detector
sends out an end-of-life trouble message. When the message is transmitted,
replace the CO sensor module.
To determine the months until end-of-life, run a maintenance report from the LCD
screen. See “Device maintenance reports” on page 52.