Multimeter Tests
Copper and G.fast/VDSL2/ADSL2+ Multi-play Test Set 135
Isolation
Selected Pair allows you to select All Pairs or a single wire pair for
measurement.
Soak Voltage allows you to apply a voltage value to the pair(s)
currently under test. Select a value from the list or User Defined to
enter your own value. If your unit has the HIVOLT (high voltage)
software option configured, the Soak Voltage parameter values are
extended up to 500 V max.
Note: Since Soak Voltage is inherited in other tests that use the Isolation test, the
HIVOLT software option also applies to the following tests: Pair Detective,
User Auto Test.
User Defined Voltage (T-R/A-B) allows you to enter a voltage value
between 50 and 125 V (50 and 500 V if HIVOLT) for Tip to Ring wire
Pair.
User Defined Voltage (T, R-G/A, B-E) allows you to enter a voltage
value between 50 and 125 V (50 and 500 V if HIVOLT) for Tip or Ring to
Ground wire Pair.
Soak Time sets a soak period of time from 1 to 60 seconds.
Threshold Settings button opens a new page that allows you to set
isolation thresholds for each pair.
Result displays a snapshot of the pass/fail status, Resistance and a
Soak Counter for each selected wire pair. When running the test, the
counter counts to the selected Soak Time by the second, for the wire
pair(s) under test.
Test Status displays the current state of measurement or an error
message.
To select the parameter values:
1. Press the up/down arrow keys to highlight the desired parameter.
2. Press the left/right arrow keys to view and select the options.