Minimum Height of Tabletop: 500mm (Allowable Deviation: ±10mm)
Tabletop Travel Range:
Horizontal Lateral: 280mm ±10mm
Horizontal Longitudinal: 680mm ±10mm
Standard Table (GCTBL-C6)
Height from Tabletop to floor: 525mm~825mm (Allowable Deviation: ±10mm)
Table Detector Travel Range: 510mm (Allowable Deviation: ±10mm)
Minimum Height of Tabletop: 525mm (Allowable Deviation: ±10mm)
Tabletop Travel Range:
Horizontal Lateral: 220mm ±10mm
Horizontal Longitudinal: 990mm ±10mm
Digital Wallstand (5181666/5136848-2/GCWS-C1/GCEWS-C1/NBS2100)
Digital Detector:
Detector Vertical Travel Range along the Wallstand Column: 1500mm ±10mm
Detector Negative Rotation Range: -20° ~0° (Allowable Deviation: ±2°)
Detector Positive Rotation Range: 0°~90° (Allowable Deviation: ±2°)
Manual Wallstand (GCMWS-C6)
Digital Detector:
Detector Vertical Travel Range along the Wallstand Column: 1500mm ±10mm
OTS:
OTS Column Vertical Travel Range: 1500mm (Allowable Deviation: ±10mm)
OTS Column Rotation Range: -180° ~ +180° (Allowable Deviation: ±2°)
X-Ray Tube Rotation Angle around the lateral arm: -180° ~ +135° (Allowable Deviation: ±2°)
Collimator Rotation Angle around the vertical arm: -90° ~ +90° (Allowable Deviation: ±2°)
Lateral Travel Range: 856mm
Longitudinal Travel Range: 2805mm
Braking Force
Digital Table: The braking force of the linear movement 100N
Standard Table: The braking force of the linear movement 100N