GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A001 - AWS Troubleshooting - No Display on Power Up
Page no. 1013 Chapter 9
JC-TSG-A-001.fm
1-4 Monitor Video Signal Checks
From page 1012
No display on monitor, but monitor
operation LED is lit
Reseat the RAM.
If the problem persists, start again
on page 1010, or ask for help from
your OLC.
Yes
Yes
Yes
Yes
Check ADS Computer for
video signal problems
Check the video
cable. Is it
defective?
No
No
RAM badly seated?
No
CPU badly seated?
No
Defective Graphics
card or Motherboard
video chip?
No
Check the monitor
synchronization
cables. Are they
defective or wrongly
connected?
Yes
Change the video cable.
Interchange the H sync. and V
sync. cables on the monitor BNC
inputs, or change the cables.
Reseat the CPU.
Change the ADS Computer