Chapter 9 Page no. 1024
JC-TSG-A-010.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A010 - Troubleshooting Bad Pixel Calibration Failure
3-2-3-1 Corrective Steps if Bad Pixels are Close to the Detector Edges
If you have seen any white band on the border(s) of a flat field image, use the troubleshooting flowchart
in the Illustration 3, page 1025.
3-2-3-2 Corrective Steps if no Bad pixel is Close to the Detector Edges
If the collimator blades check fail, use the service escalation process.