GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card TSG A010 - Troubleshooting Bad Pixel Calibration Failure
Page no. 1023 Chapter 9
JC-TSG-A-010.fm
ILLUSTRATION 2 - DECISION TREE FOR BAD PIXEL CALIBRATION FAILURE
3-2-1 Load Detector Parameters
Check on the Service Desktop home page that the detector ID is the same as the one on the detector
CD-ROM or floppy disk. If any doubt, also check that the actual labeling on the detector matches the
detector ID written on the detector CD-ROM or floppy disk.
If the detector IDs are the same, return to the troubleshooting decision tree in Illustration 2, page
1023 and continue with the other troubleshooting steps.
If the detector IDs are not the same, reload the detector parameters according to Load Detector
Parameters on page 1103 in Job Card D/R A141 - Image Receptor/Digital Detector, then return to
the troubleshooting decision tree in Illustration 2, page 1023 and continue with the other trouble-
shooting steps.
3-2-2 Setup Checks
For bad pixel calibration, it is important that:
The flat field phantom used in this procedure is very clean and free from imperfections
All surfaces in the beam (detector, tube, filter, etc.) are very clean and free from imperfections, which
can affect calibration.
The flat field phantom is well positioned on the detector: centered laterally, well covered on each
detector edge.
3-2-3 Collimator Blades Check
If you see a white band on the border(s) of a flat field image, it may be due to the shadow of the collima-
tor blade or the collimator blade itself.
Load Detector Parameters (see section 3-2-1,
Load Detector Parameters on page 1023)
Perform Bad Pixel Calibration (see
Job Card CAL
A041 - Bad Pixel Calibration
on page 1693)
Setup checks (see section 3-2-2, Setup Checks on
page 1023)
Perform Bad Pixel Calibration (see
Job Card CAL
A041 - Bad Pixel Calibration
on page 1693)
Bad Pixel calibration fails
NO
Bad Pixel calibra-
tion passes?
Perform post-corrective ac-
tions (see
section 4-1, Cali-
bration to be performed after
bad pixel calibration
on page
1026).
Detector ID of CSD = Detector ID of
Detector CD-ROM or Floppy Disk?
Bad Pixel calibra-
tion passes?
YES
.End
Collimator blades check (see section 3-2-3, Collimator Blades
Check
on page 1023)
NO
YES
YES
NO