Chapter 3 Page no. 104
S2100D How To.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Senographe Operations - How to...
ILLUSTRATION 3 -
TUBE CALIBRATION
CTUBEHTR_D
LF/MO SF/MO
HEATER WarmUpCTubH
CALIB
LF/RH
PARAM
SF/RH
HEATER: Select Foc/Track
HEATER: LF/MO CALIB
CALIB PARAM
HEATER: SF/MO CALIB
CALIB PARAM
HEATER: LF/Rh CALIB
CALIB PARAM
HEATER: SF/Rh CALIB
HEATER: LF/MO CALIB
Currents LF/MO (28 values kV/A)
I_VALUE
CHANGE
C_GAIN
Cathode Gain LF/MO GK_Ax=(3 values)
NEXT
Modif LF/MO cathode gain GK_Ax=(value)
VALID
CHANGE
Currents LF/MO modif (value) (value)
NEXT VALID
WARNING
THE TWO MENUS CTUBE AND CTUBH
ARE NOT INTENDED FOR SERVICE USE.
Same menu
structure as for
LF/MO
A
B
C