GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Senographe Operations - How to...
Page no. 105 Chapter 3
S2100D How To.fm
TUBE WARM-UP x/y kV
HTR_DELTA Parameters
LF/MO SF/MO LF/RH SF/RH
GENERATOR CALIBRATION
BIASTUBE HTR_SCL mA_measKV
HTR_SCL HARDWARE CALIB
CALIB
TRACK2TRACK1
PARAM
LF/Mo HTR_DELTA display DEL_x=(3 values)
Same menu
structure as for
LF/MO
TRACK1
HTR_SCL HARDWARE TRACK 1
1st pt 2nd pt calcul
1st pt
HTR_SCL CALIB TRACK1 PT1 ref=(value)
CMDE CALIB Meas
HTR_SCL TRACK1 PT1 modif ref=(value)
NEXT VALID
HTR_SCL TRACK1 1st point meas=(value)
NEXT VALID
VALID
PARAM HTR_SCL track 1 x=(2 values)
CHANGE
PARAM HTR_SCL modif track 1 x=(value)
NEXT VALID
THIS MENU IS ONLY
ACCESSIBLE WHEN BOTH
FILAMENT CURRENT VALUES
HAVE BEEN MEASURED AND
ENTERED VIA THE 1ST PT AND
2ND PT MENUS.
THE MENU ICH_M, WHICH IS USED TO
ENTER THE MEASURED FILAMENT
CURRENT VALUE FOUND IN THE CALIB
MENU, IS ONLY ACCESSIBLE AFTER USE
OF THE CALIB MENU.
A
B
HTR_SCL HARDWARE TRACK 1
GENE
TUBE HTR_SCL
A
B
C
Ill. 4
Same menu
structure as for
Same menu
structure as for