GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Senographe Operations - How to...
Page no. 107 Chapter 3
S2100D How To.fm
PRESSING THE "MEASUR" KEY UP-
DATES THE DISPLAYED VALUE OF THE
THEORETICAL MEASURED X-RAY
TUBE CURRENT VALUE
GENERATOR CALIBRATION
BIASTUBE HTR_SCL mA_measKV
kV CALIBRATION
CALIB PARAM
kV CALIBRATION
1st pt calcul2nd pt
kV CALIB 1st point ref=(value) kV
kV CALIB 1st point
kV REFERENCE 1st point ref=(value) kV
CALIBREF kV_M
NEXT VALID
MEASURED kV 1st meas=(value)
NEXT VALID
Same menu
structure as for
1st pt
VALID
THIS MENU IS ONLY ACCESSIBLE
WHEN BOTH HV VALUES HAVE
BEEN MEASURED AND ENTERED
VIA THE "1ST PT" AND "2ND PT"
MENUS.
kV PARAMETERS x=(2 values)
CHANGE
PARAM kV MODIF x=(value)
NEXT VALID
mA MEASUREMENT
CALIB PARAM
mA MEASURE CALIBRATION
CALIB I_MEAS CALCUL
mA MEASURE CALIBRATION meas=x mA
MEASUR
MEASURED CURRENT meas=(value)
NEXT VALID
mA MEASUREMENT CALC
VALID
mA MEASUREMENT PARAM f/mA=(value)
CHANGE
mA MEAS PARAM MODIF f/mA=(value)
NEXT VALID
THIS MENU IS ONLY
ACCESSIBLE WHEN X-RAY
TUBE CURRENT HAS BEEN
MEASURED AND ENTERED
VIA THE "CALIB" AND
"I_MEAS" MENUS.
A
B
THIS MENU IS ONLY
ACCESSIBLE WHEN 1ST PT
KV VALUE HAS BEEN
MEASURED VIA THE "CALIB"
MENU.
GENE
BIAS mA_measkV
A
Ill. 3

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals