Chapter 9 Page no. 1224
JC-DR-A-221.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A221 - Rotation Board
6 PROCEDURE
6-1 Disassembly of the Rotation Board
1. Release and remove the six screws (2) securing
the rotation board protection plate (3) (2 mm allen
wrench).
2. Disconnect all cables from the Rotation Board (4).
3. Remove the six hex nuts (2) securing the rotation
board (5.5 mm open-ended wrench)
4. Remove the Rotation Board.
6-2 Reassembly of the Rotation Board
• Reverse the disassembly procedure.
Add the following step to the reassembly procedure:
- Apply blue Loctite on the threads of the six screws securing the rotation board supporting plate
and tighten them with the Torque wrench (0.8 Nm +/- 0.1 Nm).
CAUTION
When reconnecting the cables to the Rotation Board, ensure that you route the W215
cable so it does not pass over Rotation Board. This precaution avoids potential interfer-
ence caused by component electronic noise.
2
3
4

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals