GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A221 - Rotation Board
Page no. 1225 Chapter 9
JC-DR-A-221.fm
7 COMPLETION
7-1 Firmware Check
Check whether the new Rotation Board firmware needs to be updated, as follows:
1. Reboot the Senographe system.
2. Go to the Service Desktop home page (see Launching the Service Desktop on page 120).
3. Consult Table 1 on page 63 in the section Component FW/SW to ADS Version compatibility Matrix
on page 63. Determine the system level from the ADS/IDC/Generator CPU versions displayed on the
Service Desktop home page, and by consulting Table 1 on page 63. The system level will be defined
by one of the System Level (Model Number) columns in Table 1 on page 63.
4. Check whether the current firmware version(s) for the new Rotation Board corresponds to the
required firmware version for the that system level.
If the current firmware version(s) match the required version(s) quoted in Table 1 on page 63, no
additional steps are required.
If the current firmware version(s) does not match the required version(s), you must perform a firm-
ware update of the mismatched firmware version(s) so that the new Rotation Board firmware ver-
sion(s) match those quoted in Table 1 on page 63. To update the new Rotation Board firmware
follow the instructions supplied in the Gantry Software Update CD-ROM that is kept with the sys-
tem.
7-2 Calibrations
Carry out the calibration procedure: Job Card CAL A021 - Arm Rotation Calibration on page 1679.
7-3 Check Rotation Movement
1. Move the compression paddle out of compression. Check that all four sets of arm left/right control
buttons are operational (check each button for rotation movement of column).
2. Rotate column to its left and right limits. Software preset limits must stop movement (185° counter-
clockwise, 165° clockwise).