Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A288 - Collimator Lamp
Page no. 1385 Chapter 9
JC-DR-A-288.fm
6. Disconnect the lamp connector (8).
7. Release the two screws (9) securing the lamp (10) and its support (2.5 mm allen wrench). Be careful
not to damage or move the mirror when you release the screws.
8. Remove the lamp.
8
10
9